BPIT, Inc. - Model T200P Photolytic True NO2/NO/NOx Analyzer with NumaView™ Software

Model T200P Photolytic True NO2/NO/NOx Analyzer with NumaView™ Software

By: Teledyne Instruments API


The Model T200P provides measurements of NO and NO2 using our Model T200 NOx analyzer combined with a patented high efficiency photolytic converter. Even low temperature molybdenum converters transform other nitrogen-containing compounds such as HNO3, PAN, etc. to a considerable extent. Simultaneous measurements of NO2 performed with molybdenum and photolytic converters have shown significantly different results in the presence of such compounds.


BPIT, Inc. - Quotation icon Get a quote »



Features

  • Ranges: 0-50 ppb to 0-4,000 ppb, user selectable
  • High efficiency photolytic converter with specific conversion of NO2
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • 0.40 ppb lower detectable limit
  • Catalytic ozone scrubber
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty

Specifications:

Ranges

Min: 0 - 50 ppb full scale for NO and NO2
Max: 0 - 4,000 ppb full scale for NO2;
0-20,000 ppb full scale for NO
 (selectable, dual-range supported)
 

Measurement Units

ppb, ppm, μg/m3, mg/m3 (selectable)

Zero Noise

< 0.2 ppb (RMS)

Span Noise

< 0.5% of reading (RMS) above 50 ppb

Lower Detectable Limit

0.4 ppb

Zero Drift

< 0.5 ppb/24 hours

Span Drift

< 0.5% of full scale/24 hours

Lag Time

20 seconds

Rise/Fall Time

< 60 seconds to 95%

Linearity

1% of full scale

Precision

0.5% of reading above 50 ppb

Sample Flow Rate

500 cc/min ±10%

Power Requirements

100V-120V, 220V-240V, 50/60 Hz

Analog Output Ranges

10V, 5V, 1V, 0.1V (selectable)

Recorder Offset

±10%

Included I/O

1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs

Optional I/O

1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4-20mA current outputs

Operating Temperature Range

5 - 40ºC

Dimensions (HxWxD)

7” x 17” x 23.5” (178 x 432 x 597 mm)

Weight

Analyzer: 40 lbs (18 kg)
External pump: 15 lbs (7 kg) ‚Äč