The Model T100U analyzer achieves low level SO2 measurements using the proven UV fluorescence principle and advanced electronics. The T100U combines high sensitivity with a wide dynamic measurement range, making it ideal for ambient air quality and other low-level applications.
— With NumaView™ premium T Series software —
Features:
Specifications:
Ranges |
Min: 0-5 ppb full scale Max: 0-20,000 ppb full scale (selectable, dual ranges) |
Measurement Units |
ppb, ppm, μg/m3, mg/m3 (selectable) |
Zero Noise |
25 ppt (RMS) |
Span Noise |
0.5% of reading (RMS) above 5 ppb |
Lower Detectable Limit |
50 ppt |
Zero Drift |
< 0.2 ppb/24 hours |
Span Drift |
< 0.5% of full scale/24 hours |
Lag Time |
30 seconds |
Rise/Fall Time |
< 140 seconds to 95% |
Linearity |
1% of full scale |
Precision |
0.5% of reading |
Sample Flow Rate |
650 cm3/min ±10% |
Power Requirements |
100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges |
10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset |
±10% |
Included I/O |
1 x Ethernet: 10/100Base-T 2 x RS232 (300-115,200 baud) 2 x USB device ports 8 x opto-isolated digital outputs 6 x opto-isolated digital inputs 4 x analog outputs |
Optional I/O |
1 x USB com port 1 x RS485 8 x analog inputs (0-10V, 12-bit) 4 x digital alarm outputs Multidrop RS232 3 x 4-20mA current outputs |
Operating Temperature Range |
5 - 40ºC (with US EPA Equivalency) |
Dimensions (HxWxD) |
7” x 17” x 23.5” (178 x 432 x 597 mm) |
Weight |
45 lbs (20.5 kg) |
Certifications |
US EPA: EQSA-0495-100 |