BPIT, Inc. - LabGard ES Air NU-543 Class II, Type A2 Biological Safety Cabinet

LabGard ES Air NU-543 Class II, Type A2 Biological Safety Cabinet

By: Nuaire


LabGard Air offers the very best in personnel, product, and environmental protection through the use of airflow probe to monitor and maintain real time airflow and displays airflow speeds on the OLED interface. The ability to choose up to three (3) different access opening heights allows our customers the option to select a larger access opening to improve technician movement and reach; select a smaller access opening to lower your total cost of ownership; or select somewhere in-between to get the best of both heights.


BPIT, Inc. - Quotation icon Get a quote »



Features:

Technology:

FlowGard™ Electronic Control System

FlowGard™ ECS offers a user friendly design with rich functionality. Basic cabinet functionality (Motor Blower, Lights, UV Light, Outlet(s), Alarm Silence) is controlled through easily identifiable buttons. Cabinet's inflow and downflow air speed is displayed on the FlowGard™ ECS OLED display screen. Airflow performance is monitored and controlled via intelliFlow™ - A Digital Pressure Monitor, the same technology used by service professional who certify your cabinet at minimum once a year.

Ultra High Efficiency Motor

A single ultra-high efficiency ECM motor with a properly sized fan blade minimizes energy usage while extending the life of filters. LabGard offers the lowest total cost of ownership compared to other models.

HEPEX Zero Leak Airflow System

All LabGard positive pressure plenums are surrounded by a vacuum relative to the room. Meaning in the event of a leak via gasket failure, all contaminated air in the plenum is still forced through HEPA filtration. The HEPEX™ system also works to provide true laminar airflow by evenly distributing plenum air over the supply HEPA filter. The flexible design eliminates any vibration transfer from the motor blower as well as creates a serviceable work environment for Certifiers after decontamination.

 

Safety is Priority Number 1:

True Laminar Airflow

Unidirectional airflow moving along parallel flow lines at a constant velocity of 60 fpm (0.30 m/s) minimizes air turbulence within the work zone lessening the chance for cross contamination.

Dynamic Air Barrier

A strong inflow air velocity of 105 fpm (0.53 m/s) over an aerodynamic airfoil creates a dynamic air barrier offering maximum personnel protection.


Specifications:

Model NU-543-300 NU-543-400 NU-543-500 NU-543-600
Nominal Width 3 Foot (0.9 m) 4 Foot (1.2 m) 5 Foot (1.5 m) 6 Foot (1.8 m)
Performance Specifications:
  • Personal Protection
  • Product Proection

NSF/ANSI 49

EN 12469
NSF/ANSI 49 Class II
Construction

All welded stainless steel 16/18 Gauge, Type 304 pressure tight design
 
Diffuser for Air Supply (Metal) Non-fammable
HEPA Filter Seal Type:
  • Supply Filter 99.99% @ 0.3 Micron
  • Exhaust Filter 99.99% @ 0.3 Micron
HEPEX Seal Neoprene, Spring Loaded
Fumigation per NIH/NSF Procedures Yes
Work Access Opening:
  • Standard Opening Height
10" (254 mm)
  • Optional Opening Height
12" (305 mm)
  • Optional Opening Height
8" (203 mm)
  • Standard Inflow Velocity
105 fpm (0.53 m/s)
Viewing Window Standard Safety Plate Sliding Glass Fully Closed to 21" (533 mm) Open
Exterior Dimensions:
Height 61.875" (1572 mm) 61.875" (1572 mm) 61.875" (1572 mm) 61.875" (1572 mm)
Height (min) 59" (1499 mm) 59" (1499 mm) 59" (1499 mm) 59" (1499 mm)
Width 41.625" (1057 mm) 53.625" (1362 mm) 65.625" (1669 mm) 77.625" (1972 mm)
Depth (Armrest Removed) 31.4375" (799 mm) 31.4375" (799 mm) 31.4375" (799 mm) 31.4375" (799 mm)
Interior Dimensions:
Height 28.5" (724 mm) 28.5" (724 mm) 28.5" (724 mm) 28.5" (724 mm)
Width 34.375" (873 mm) 46.375" (1178 mm) 58.375" (1483 mm) 70.375" (1788 mm)
Depth 25.75" (654) 25.75" (654) 25.75" (654) 25.75" (654)
Sound Pressure Level per ISO 4871 Not to exceed 55 dbA Not to exceed 56 dbA Not to exceed 58 dbA Not to exceed 60 dbA